Properties:
Related Categories |
Additional Standards,
Air PT,
Analytical Standards,
Analytical/Chromatography,
Proficiency Testing
|
Description:
Analysis Note
This sample will contain up to a maximum of 10 randomly selected analytes from the analyte list
General description
This proficiency testing sample is produced in accordance with ISO/IEC 17043:2010. All information regarding the use of this material can be found in the reporting packet supplied for each sample.
Other Notes
Proficiency testing samples are presented as single blind. No values for the samples are released until the testing period concludes. Values of analytes vary lot to lot.
Proficiency Testing Schedule (2017-2018)
Air
AIR17-2 April 12 - May 26, 2017
AIR17-4 October 11 - November 24, 2017
AIR18-2 April 11 - May 25, 2018
AIR18-4 October 10 - November 23, 2018
Quick-Turn Study - (non-scheduled proficiency test that is available at any time)
Preparation Note
Sample preparation
This sample is provided as a 2mL ampule and a vial containing activated charcoal.??
1) Place 100 mg of the supplied activated carbon into an appropriate extraction vial.
2) From the supplied ampule and using an air-tight syringe, add 100μL to the activated carbon in the vial.
3) Add 1mL of carbon disulfide to the vial and cap.
4) Allow to stand for 30 minutes with occasional agitation.
5) Sample is ready for analysis.
6) Report your value as μg/vial (μg in the 100μL injection).
Safety:
RIDADR
NONH for all modes of transport